AFM Tip Shape Effects

AFM Tip Shape Effects

Atomic Force Microscopy (AFM) is a measurement technique that is capable of very high magnification imaging. Unlike other techniques, AFM resolution is not limited by photon or electron wavelengths. AFM uses a super-sharp AFM tip for mechanical scanning of sample surfaces in a similar fashion to a blind man who reads Braille code with his fingers.

Images obtained by scanning a sample with an AFM tip are strongly dependent on the AFM tip geometry. AFM scans are a result of the convolution of the AFM tip and the surface. The smaller the surface features compared to the AFM tip dimensions, the less accurate the sample image on the computer screen.

AFM tip shape effect

The two most important AFM tip related factors that affect the quality of the AFM scans are the apex radius of curvature and the sidewall angles (Fig. 1). The AFM tip radius is the real limitation of the ultimate resolution that an AFM measurement can yield, while the AFM tip sidewall angles affect the accurate imaging of steep slopes.

Cross section schematic of an AFM probe
Fig. 1: Cross section schematic of an AFM
tip with important parameters.

How does the AFM image compare to the real surface topography? The main effects of AFM tip geometry are shown in Fig. 2 and explained below.

  1. High surface features appear broadened laterally. The broadening at the top of the feature is equal to the AFM tip curvature diameter and increases additionally down the sidewalls (Fig. 2 A, Fig. 3) due to the AFM tip sidewall angles (3). The wider the AFM tip, the more noticeable this effect is.
  2. Sharp feature edges appear rounded. The rounding occurs for both acute and obtuse angles (Fig. 2 B, Fig. 4). The larger the AFM tip radius of curvature, the more noticeable this effect is.
  3. Apparent feature sidewall angles are limited to the AFM tip sidewall angles. All sidewalls with steeper slopes than the sidewall angles of the tip appear to have the tip sidewall angles (Fig. 2 C). Less steep sidewalls can be measured without distortions (Fig. 2 E).
  4. Trenches appear narrower. Lateral dimensions of the trenches shrink (Fig. 2 F) as a direct consequence of the broadening of high features (1).
  5. Trench depth between adjacent structures can appear shallower due to the inability of the AFM tip to enter the narrow space in-between and touch the bottom surface (Fig. 2 G, Fig. 4). Step heights around broader trenches are measured correctly (Fig. 2 D).
Cross section schematic of a scanned surface
Fig. 2: Cross section schematic of a scanned surface (above) and its image (below) illustrating different image distortion effects: lateral broadening (A), edge rounding (B), slope reduction (C), trench narrowing (F) and trench depth reduction (G).
AFM topography scan (left) and cross section (right)
Fig. 3: AFM topography scan (left) and cross section (right) of a TMV virus. Notice the lateral broadening and the side slopes of the actually circular cross section of the virus
Cross sections of AFM topography scans
Fig. 4: Cross sections of AFM topography scans taken on the TipCheck with a sharper AFM tip (above) and a duller AFM tip (below) at the same vertical and horizontal scale. Notice the additional edge rounding and the decreased heights when scanning with a duller AFM tip

As frictional wear dulls the AFM tip progressively, the above described effects become more and more pronounced. After a certain point, the contact area of the AFM tip with the surface is so large that it dominates the shape of small features and all of them appear to have very similar shape and the same orientation (Fig. 5, Fig. 6). This shape is actually the AFM tip apex shape rotated by 180 degrees and its geometry depends on the cross section of the AFM tip pyramid.

Horizontal cross section schematic
Fig. 5: Horizontal cross section schematic illustrating feature shape distortion when scanning small features with a dull AFM tip.
AFM topography scan of the TipCheck
Fig. 6: AFM topography scan of the TipCheck with a dull AFM tip illustrating
feature shape distortion (tip imaging).

When the AFM tip is broken, for example as a result of improper engage with the surface, and there are additional non-idealities on the apex, they can show up as fake feature details in the AFM image (Fig. 7, Fig. 8).

Cross section schematic
Fig. 7: Cross section schematic illustrating fake feature imaging as a
result of a broken AFM tip.
AFM topography (left) and deflection (right) scans
Fig. 8: AFM topography (left) and deflection (right) scans of TMV viruses with a broken AFM tip illustrating fake details. Notice the 3 steps on the bottom and the 2 steps on the top of the TMV viruses with actually circular cross section.

The artifacts can be even more severe. When the pyramid has more than 1 AFM tip that can contact the sample surface during scanning, multiplication of feature images occurs (Fig. 9, Fig. 10).

Cross section schematic illustrating double imaging of sample surface
Fig. 9: Cross section schematic illustrating double imaging of sample surface features as a result of a broken AFM tip.

Some AFM tip shape effects such as shape distortions or artificial details are not hard to detect. They are present throughout the scanned image in an identically repeatable and clearly recognizable pattern. Other effects are more subtle and users can not immediately identify them just by a quick look at the computer screen. Therefore, good judgment must be used when measuring nanoscale distances or interpreting nanoscale feature shapes. By knowing the peculiarities of AFM measurements imposed by the AFM tip shape, it is nevertheless possible to extract useful information from the AFM scans or recognize a false measurement that needs to be redone.

IMPORTANT!
Scan the sample surface

Understanding AFM tip shape and its effects on AFM measurements is essential for performing successful AFM analysis.

Scan the sample surface

Choosing the right AFM probes is critical for achieving good AFM measurement results. AFM tip sharpness is directly related to the maximum achievable resolution

Scan the sample surface

All image distortions caused by dull or broken AFM tips can also appear as a result of particles sticking to the AFM tip and acting as an AFM tip apex. Many AFM systems operate outside of clean-room facilities. Keeping the working space, probes and samples clean will reduce particle contamination and improve the quality of the AFM scans.