www.nanosensors.com


Contact us at:
tel: + 49 (0) 6441 870 6272
fax: + 49 (0) 6441 870 6274
info@nanoandmore.com
The new PointProbe® Plus (PPP) combines the well-known features of the proven PointProbe® series such as high application versatility and compatibility with most commercial SPMs with a further reduced and more reproducible tip radius as well as a more defined tip shape. The excellent tip radius and the minimized variation in tip shape provide more reproducible images and enhanced resolution.
NANOSENSORS™ PPP-NCHR probes are designed for non-contact mode or tapping mode AFM (also known as: attractive or dynamic mode). This AFM probe combines high operation stability with outstanding sensitivity and fast scanning ability.
This AFM probe offers unique features:
- excellent tip radius of curvature
- highly doped to dissipate static charge
- Al coating on detector side of cantilever
- high mechanical Q-factor for high sensitivity
This product features alignment grooves on the backside of the holder chip.
| Shape | Length | Width | Thickness | Force Const. | Res. Freq. |
|---|---|---|---|---|---|
| beam | 125 µm(115 - 135 µm)* | 30 µm(22.5 - 37.5 µm)* | 4 µm(3 - 5 µm)* | 42 N/m(10 - 130 N/m)* | 330 kHz(204 - 497 kHz)* |
| Shape | Tip Radius |
|---|---|
| Standard | <7 nm( < 10 nm guaranteed ) |
* guaranteed range
The reflex coating is an approximately 30 nm thick aluminum coating on the detector side of the cantilever which enhances the reflectivity of the laser beam by a factor of about 2.5. Furthermore it prevents light from interfering within the cantilever. The virtually stress-free coating is bending the cantilever less than 3.5% of the cantilever length.
Datasheet for all AFM probes with sets of 10 or 20 probes.
Datasheet for up to 32 probes with full wafer.
Product Availability: |
From stock |
Price Description: |
Does not include VAT or customs duties. |